CE Marked Device on Great Britain Market
Can a CE-marked medical device still be placed on the Great Britain market, until when, and what does MHRA registration require?
Yes. A CE-marked device can still be placed on the Great Britain market, but only until fixed transitional end dates, and it must be registered with the MHRA before it is placed on the market. According to the MHRA's "Regulating medical devices in the UK" guidance on GOV.UK (seen as a search snippet), the end dates are:
| CE-marking basis | Acceptable on the GB market until |
|---|---|
| EU MDD / EU AIMDD (with a valid declaration and CE marking) | 30 June 2028, or earlier if the certificate expires first |
| EU IVDD | 30 June 2030, or earlier if the certificate expires first |
| EU MDR (including custom-made devices) and EU IVDR | 30 June 2030 |
Possible change ahead. The MHRA ran a consultation from 16 February to 10 April 2026 on three proposals:
- Indefinite recognition of devices compliant with the EU MDR and EU IVDR.
- Extending the transition for MDD-certified devices so it lines up with the EU's own MDR transition.
- An international reliance route for CE-marked devices that would fall into a higher risk class under GB rules.
I found no official government response or amending legislation yet, so the dates above remain the ones in force.
Expired MDD/AIMDD or IVDD certificates. Certificates kept valid by the EU's revised transitional arrangements (March 2023 for medical devices, June 2024 for IVDs) can be relied on in GB. To register on that basis you must upload one of the following:
- A notified body letter confirming it has received the application for conformity assessment and that a written agreement is in place.
- If you have no such letter, the MHRA's completed Article 120 (MDR) or Article 110 (IVDR) extension confirmation template.
What MHRA registration requires (UK MDR 2002, GB market)
-
Who must register: anyone who sells, leases, lends or gifts any of the following, and the MHRA must be told before the device is placed on the market:
- Class I, IIa, IIb or III devices they manufactured, or refurbished or re-labelled under their own name
- systems or procedure packs containing at least one device
- custom-made devices
- IVDs, including those undergoing performance evaluation
-
Timing: registration is needed before the first supply of each device on the GB market (not each type or model), whether supplied free or for payment [4]. A device that is only put into service, such as one a hospital makes and uses in-house, doesn't have to be registered; registration is encouraged [5].
-
Manufacturers outside the UK must appoint a single UK Responsible Person, who handles registration for all of that manufacturer's devices. DORS accounts of former GB Authorised Representatives that had not switched to the UK Responsible Person role by 30 March 2026 have been closed [3].
-
Manufacturer data required:
- legal entity name and address as shown on the labelling
- company type
- administrative contacts
- where applicable, a letter of designation: a legal contract naming the exclusive UK Responsible Person and the mandatory tasks it carries out for the manufacturer
-
Device data required:
- which legislation applies, and the device class
- GMDN code and term
- Basic UDI-DI and UDI-DI (if applicable)
- brand or trade name, model or version, and catalogue number
- UK Approved Body or EU Notified Body, where applicable
- attributes such as sterility, latex content and MRI compatibility
- copies of conformity assessment certificates or self-certification declarations
The MHRA may ask for further technical documentation before confirming registration [6].
-
Importers: a GB importer must tell the manufacturer or UK Responsible Person, who then submits the importer's details to the MHRA [7].
-
System: you apply through the Device Online Registration System (DORS). You are not registered until the MHRA confirms it. Registration is not an approval, so you may not claim MHRA endorsement or use MHRA logos [8]. Registered details appear on the public register, PARD [9].
-
Fees: since 1 April 2026 an annual fee replaces the old one-off registration fee. For 1 April 2026 to 31 March 2027 it is £300 per Level 2 GMDN category (or Level 1 where no Level 2 exists), worked out from the account as it stands on 31 March and charged pro rata for registrations made during the year [10].
-
Keeping it up to date: changes must be reported to the MHRA. Some changes need a new registration and may carry a fee: changing the legislation (for example from MDD to UK MDR or EU MDR), changing the UK Responsible Person, or changing device characteristics such as class, GMDN or trade name [11][12]. The registration must stay in place for as long as the device is made available in GB [13].
-
EU MDD/AIMDD custom-made devices can no longer be placed on the GB market. They can be registered under UK MDR 2002 Part II or Part III instead [6].
Northern Ireland is separate. Since 28 May 2026, devices other than custom-made ones are registered in EUDAMED, and MHRA registration is not required for them [14].
Approach: I read the MHRA guidance "Register medical devices to place on the market". The CE-acceptance dates and the consultation status come from official GOV.UK pages, seen as search snippets.