Q.01

How do I register a medical device with the MHRA?

You register through the MHRA's online Device Online Registration System (DORS), and you must do it before you place the device on the Great Britain market. First you create a DORS account. Once the MHRA accepts it, you submit the manufacturer and device details along with your conformity documents, and you pay an annual fee for each GMDN category. If the manufacturer is outside the UK, a UK Responsible Person has to do the registration. Northern Ireland works differently: since 28 May 2026 most devices are registered on EUDAMED instead.

Approach: I read the MHRA's current GOV.UK guidance "Register medical devices to place on the market". All requirements below come from it.

1. Check that you need to register (Great Britain)

  • The UK MDR 2002 requires you to tell the MHRA before you place a device on the GB market [1].
  • You must register if you sell, lease, lend or gift any of the following [1]:
    • Class I, IIa, IIb or III devices you made, refurbished or re-labelled under your own name
    • systems or procedure packs
    • custom-made devices
    • IVDs, including IVDs undergoing performance evaluation
  • If a device is only put into service (for example, a hospital making and using its own devices), registration is encouraged but not required [2].
  • Confirm that the product meets the Regulation 2 definition of a medical device or IVD. If it's a borderline case, the MHRA points you to its borderline products guidance [3].

2. Decide who registers

  • A manufacturer based outside the UK must appoint a single UK Responsible Person for all of its devices. That person takes over the registration duty [1].
  • The UK Responsible Person has to upload a letter of designation. This is a legal contract naming them as the exclusive UK Responsible Person and listing the mandatory tasks [3].
  • Importers into GB must tell the manufacturer or UK Responsible Person, who then submits the importer's details to the MHRA. Distributors that do not import don't need to register [4].

3. Create a DORS account, then register the devices

  • The MHRA emails you to say whether your account request is accepted. You only count as registered once you have entered the device details and received confirmation. The MHRA may ask for technical documentation before it confirms [5].

  • Registration is not an approval or endorsement. You may not claim that it is or use MHRA logos [5].

  • Manufacturer data required:

    • legal entity name and address as shown on the label
    • company type
    • administrative contacts (up to 15)
    • the UK Responsible Person's letter of designation, where applicable

    [3]

  • Device data required:

    • the legislation that applies, and the device class
    • GMDN code and term (you can pick the term inside DORS without joining the GMDN Agency)
    • Basic UDI-DI and UDI-DI, where applicable
    • brand or trade name, model or version, and catalogue or reference number
    • the UK Approved Body or EU Notified Body, where applicable
    • attributes such as sterility, latex content and MRI compatibility
    • copies of conformity assessment certificates or self-certification declarations

    [3]

  • Special cases:

    • Systems and procedure packs: register each one by GMDN, with at least one underlying product and a list of possible components [3].
    • IVDs in performance evaluation: register them by the time the study starts, with a Declaration for Performance Evaluation [6].
    • Coronavirus antigen and molecular tests: these need coronavirus test device approval (CTDA) before the MHRA will accept the registration [7].
    • Devices relying on extended EU MDR or IVDR CE certificates: upload the notified body letter or the MHRA's Article 120 or Article 110 extension confirmation template [8][9].
  • One application can cover up to 100 GMDN devices and up to 20,000 products in total [10].

4. Pay the fee

  • Since 1 April 2026, the old one-off registration fee has been replaced by an annual fee charged per Level 2 GMDN category. Where a category has no Level 2, the Level 1 category is used [10].
  • For 1 April 2026 to 31 March 2027 the fee is £300 per category per year. Registrations made during the year are charged pro rata [10].
  • Pay only through DORS. If you pay by BACS or CHAPS, quote the reference number DORS gives you. The deadline for 2026/27 annual fees was 6 November 2026, and accounts still unpaid after that date are suspended [10].

5. After registration

  • Your company and devices are listed on the Public Access Registration Database (PARD). IVDs in performance evaluation are not listed [11].

  • You are legally required to tell the MHRA about changes. The guidance cites regulations 7A and 19 for general devices, 21A and 21B for active implantables, and 33A and 44 for IVDs. The "Renew registration" function was removed on 1 April 2026, but you still need to review your registration regularly [12].

  • Some changes trigger a fee, for example:

    • a new GMDN category
    • a change of device class or legislation
    • a change of UK Responsible Person

    Address and contact updates, and adding products to devices already registered, are free [13].

  • If you stop supplying a device, unregister it. Post-market surveillance and vigilance duties still apply to devices already on the market [14].

Northern Ireland

  • Since 28 May 2026, all devices except custom-made ones must be registered on EUDAMED before they go on the NI market, and MHRA registration is no longer required for them [15].
  • Custom-made devices must still be registered with the MHRA within 28 days of being made available in NI [15].
  • GB manufacturers need an Authorised Representative based in the EU or NI [15].

For questions about registration, the MHRA's contact is info@mhra.gov.uk. For questions about whether a product is a device or which class it falls in, email devices.borderlines@mhra.gov.uk [16].

Q.02

What does registration cost?

For devices placed on the Great Britain market, registration now costs £300 a year for each Level 2 GMDN category your registered devices fall into (or the Level 1 category where no Level 2 exists). That rate applies from 1 April 2026 to 31 March 2027, and the MHRA says it "will change annually" [10]. The fee depends on the number of categories, not on how many devices or products you register. A new registration made part-way through the year is charged pro rata to 31 March.

Approach: I used the MHRA's GOV.UK guidance "Register medical devices to place on the market" and its DORS Fees Guidance for the fee structure that started on 1 April 2026.

How the fee is worked out

  • Charge per category. The annual fee replaced the old one-off registration fee on 1 April 2026. Each category is charged once a year, however many devices and products sit in it. A device that falls into more than one Level 2 category is also charged only once [10].
  • Lowest possible count. The MHRA checks all the Level 2 categories on your account and charges you for as few as it can. Where categories overlap, it charges under the one with the most GMDN entries [17].
  • Annual calculation. The annual fee is calculated on 1 April from the categories registered on 31 March. Categories are then fixed until the next 31 March, so a GMDN Agency reassignment during the year doesn't trigger a second charge [18][19].
  • Estimate in DORS. A calculator on your DORS Summary page shows an estimate of the annual fee [18].
  • No reductions. There are no waivers, SME rates, instalments or payment plans [17].

Registering part-way through the year

  • Pro rata method. DORS divides the annual fee by 365 and multiplies by the days left until 31 March, for each new chargeable category. You pay when you apply. The MHRA's own example: four categories registered on 2 April (364 days left) cost £1,196.70 [20].
  • Applying today. Using the same method, one new category registered today (19 November 2026, 133 days left to 31 March 2027) works out at about £109.32. That is my arithmetic, not an MHRA figure.
  • Categories you already pay for. Adding a device to a category you have already paid for costs nothing. If you unregister the only device in a category and later register a new one there, you pay the pro rata fee again. The MHRA's advice is to register the new device first, then unregister the old one [20].
  • Put into service only. Devices registered as put into service only are free [21].

Changes to a registration

  • Changes that may carry the statutory fee [13]:

    • adding devices in a new Level 2 category
    • changing device characteristics that need re-registration, where this puts the device in a new category
    • a change of UK Responsible Person
    • adding new represented manufacturers
    • changing an IVD's status, or the legislation a device is registered under. These are free if the GMDN category stays the same.
  • Free changes:

    • address, company name and contact details
    • adding products to devices already registered
    • removing devices
    • updating obsolete GMDN codes
    • adding or deactivating importers

    [13]

  • No refunds. Nothing is refunded when you unregister a device or an account [22].

Paying

  • Via DORS only. Pay through DORS. If you pay by BACS or CHAPS, use the DORS reference number. Outstanding 2026/27 fees for devices placed on the market were due by 6 November 2026, and accounts still unpaid after that date are suspended [10].
  • Closed accounts. If your renewal was due by 31 March 2026 and you didn't renew, your account has been closed. You will have to re-register and pay the new fee [23].

Northern Ireland

  • 2026/27. For NI manufacturers and those represented by an NI Authorised Representative:

    • custom-made devices pay the full fee
    • other devices pay only a partial fee, up to 27 May 2026

    [24]

  • Voluntary registrations. Non-custom-made devices that you choose to keep on, or add to, the MHRA register are charged the standard pro rata fee in 2026/27 and the full fee from 1 April 2027 [24][25].

The guidance does not give a rate for 2027/28 beyond saying the fee will change annually.

Q.03

How is Northern Ireland different?

Northern Ireland follows the EU medical device rules under the Windsor Framework, so registration there works differently from Great Britain. Since 28 May 2026, every device except custom-made devices must be registered on EUDAMED before it goes on the NI market, and MHRA registration is no longer required for those devices [15]. Custom-made devices are the only ones that still have to be registered with the MHRA, within 28 days of being made available in NI [15].

Approach: I used the MHRA's GOV.UK guidance "Register medical devices to place on the market" and its supporting documents: the Fees Guidance, the placing-on-the-market / putting-into-service guidance, and the DORS reference guides.

Legal basis

  • NI: The EU MDR (2017/745) has applied in Northern Ireland since 26 May 2021, and the EU IVDR (2017/746) since 26 May 2022 [26]. Registration of custom-made devices is governed by the Medical Devices (Northern Ireland Protocol) Regulations 2021 [15].
  • GB (England, Scotland and Wales): The UK MDR 2002 requires every device to be registered with the MHRA before it is placed on the market [27].

Where you register

Device typeGreat BritainNorthern Ireland
Non-custom-made devices (including IVDs)MHRA, through DORS, before placing on the marketEUDAMED, before placing on the market (since 28 May 2026). MHRA registration not required
Custom-made devicesMHRA, through DORSMHRA, within 28 days of being made available on the NI market
ImportersThe manufacturer or UK Responsible Person (UKRP) submits the importer's details to the MHRAEU MDR/IVDR importer obligations apply. Importers must register on EUDAMED from 28 May 2026

Sources: [15][4][28]

  • One account for both markets: UK manufacturers need only one DORS account to register in both GB and NI [15].
  • Market status of each product: In DORS you record each product as "On the GB & NI market", "On the NI market" or "On the GB market" [29].

Who registers

  • NI uses an Authorised Representative, not a UKRP.
    • Manufacturers based outside both NI and the EU must appoint an Authorised Representative based in the EU or NI.
    • This includes GB manufacturers [15].
  • Custom-made devices: who files with the MHRA.
    • If the Authorised Representative is based in NI, it "may" register them for the manufacturer.
    • If it is based in the EU, the manufacturer "must" register them itself [15].
  • Non-UK manufacturers without an NI Authorised Representative: email Device.Registrations@mhra.gov.uk for instructions on registering custom-made devices for NI [15].
  • Mandate instead of letter of designation: For NI registrations, the manufacturer or the NI Authorised Representative must submit a written mandate that meets the EU MDR and/or IVDR requirements. In GB the equivalent document is the UKRP letter of designation [3].

Custom-made device content

  • Standard to meet: In NI, custom-made devices must comply with the EU MDR [3].
  • Pre-26 May 2021 exception: Custom-made devices made under the EU MDD or AIMDD and placed on the market in an EU member state before 26 May 2021 can still be registered for the NI market only [3][30].
  • Contrast with GB: In GB you register custom-made devices under UK MDR 2002 Part II or III [3].
  • CE UKNI option: Do not choose the CE UKNI-MDR/IVDR option for self-certification. Only a UK Notified Body can carry out that type of assessment [30].
  • Definitions that apply: NI uses the EU MDR/IVDR definitions of "placing on" and "making available on" the market. A custom-made device that is only put into service is still encouraged to be recorded in DORS, and no fee applies to it [28].

Fees for NI-based manufacturers and represented manufacturers of NI Authorised Representatives

PeriodCustom-made devicesNon-custom-made devices
Annual fee charged on 1 April 2026Full feePartial fee, up to 27 May 2026
In-year pro rata registrationsStandard pro rata to 31 March 2027Partial pro rata to 27 May 2026; standard pro rata to 31 March 2027 if registered voluntarily from 28 May 2026
From 1 April 2027Full fee (registration mandatory)Full fee for any device that stays registered voluntarily

Sources: [24][25]

  • Practical point: Non-custom-made devices that stay on your MHRA account will be charged the full fee from 1 April 2027, even though registering them is now voluntary. Unregister any you don't want to keep listed.
  • Who pays: NI Authorised Representatives are among the organisations allowed to pay. They collect fees from the manufacturers they represent and pay through DORS [31].

These MHRA sources don't describe how EUDAMED registration itself works. That falls under the EU MDR/IVDR and the European Commission's EUDAMED guidance.

Q.04